Dear Carol Anne,
<p>Here are my overheads for the minutes of the Stockholm SDAST meeting.
<p>1) 041107_anod1.ps: diagram over JEM-X1 anodes on 041107 (red) compared
with the same on 040911. this does not show any wild changes, but just
a small widening of the new "death valley" around RAWX=185.
<p>2) 041101_0_anod1.ps: Almost the same as previous but for a longer period
(6 months) showing better the strong widening deepening of the "death valley",
as well as the quite new dead node at RAWX=73, and the deepening of the
old dead anode at RAWX=220.
<p>3) Anodes_list.txt: Anodes status for both JEM-X instruments.
<p>4) JMX1_Xe_time.ps: Shows the evolution of the measured position of
the Xenon line from 1335 JEM-X1 detector's spectra (kindly supplied by
PK) up to the revolution 245. (I did not shown this overhead in Stockholm
but its equivalent up to rev. 229). A quicker rising of the Xenon line
position is obviously observed since the beginning of July, probably related
to a problem with one of the calibration sources.
<p>5) JMX1_Xe_width.ps: Corresponding evolution of the detector resolution
(absolute E in red, relative in green) for the Xenon line. The relative
resolution rises up to 13% in average at the end of the period.
<pre>--&nbsp;
Dr. J&eacute;r&ocirc;me CHENEVEZ&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Danmarks Rumcenter
Danish Space Research Institute&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Juliane Maries Vej 30
<A HREF="http://www.dsri.dk/~jerome">http://www.dsri.dk/~jerome</A>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; DK-2100 Copenhagen &Oslash;
Tel. +45 35 32 57 03&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; FAX: +45 35 36 24 75</pre>
&nbsp;</html>