Gain Notes for Revolution 1519 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. To get optimal results with this revolution it is necessary to use the IC gain history. See notes in revolutions prior to revolution 1433 to find out how to do this if you don't already know. The offline gain correction is based on the two Fe sources in the JEM-X1 instrument because the Cd sources in both instruments are now too weak to use. Latest OSA software (version 10 and higher) automatically cuts out the first Science windows of data from each revolution to ensure that people do not use data from the instrument-settling period. This process can be overruled, but only by experienced users. JEM-X1: The Xe line analysis performed using the IC table shows good gain correction with all science windows (taken 2 at a time) having a Xe level within 3-4% of the ideal, with no exceptions. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications (automatically removed by OSA 10.0). JEM-X2: The Xe line analysis performed using the IC table shows good gain correction with all science windows (taken 2 SCWs at a time) having a Xe level within 3-4% of the ideal, with no exceptions. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications (automatically removed by OSA 10.0). CAO 24/03/2015