Gain Notes for Revolution 1454 Gain Notes for Revolution 1454 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. NB!!!!! This revolution has 3 separate switch-ons, each of which requires a settling period during which the energy calibration is only partially understood. This means that data taken within a couple of hours of a switch on (immediately after radiation belt exit, or manually commanded after ground station outage), will be unsuitable for energy sensitive applications. To get optimal results with this revolution it is necessary to use the IC gain history. See notes in revolutions prior to revolution 1433 to find out how to do this if you don't already know. Latest OSA software (version 10 and higher) automatically cuts out the first Science windows of data from each revolution to ensure that people do not use data from the instrument-settling period. This process can be overruled, but only by experienced users. JEM-X1: The Xe line analysis performed using the IC table shows acceptable gain correction with all science windows with a Xe peak fit that has a realistic fitted width, having a Xe level within 3-4% of the ideal. Very bad periods with poor Xe determination cluster around the switch ons, and cannot be improved with offline corretions. As usual, it is recommended that users avoid the first few science windows for energy-sensitive applications (automatically removed by OSA 10.0). Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. CAO 29/9/2014 JEM-X2: The Xe line analysis performed using the IC table shows excellent gain correction with all science windows (taken 1 at a time) having a Xe level within 2% of the ideal, except for the first few SCWs and the very last SCW. As usual, it is recommended that users avoid the first few science windows for energy-sensitive applications (automatically removed by OSA 10.0). As with JEM-X1, source 4 is no longer used for energy correction purposes due to a broken anode strip CAO 18/9/2014