Gain Notes for Revolution 1386 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. NB: Something very odd, neither temperature nor countrate related, happened to cause a sudden gain suppression on this instrument. However most of this problem has been cleared up with the use of the IC tables. To get optimal results with this revolution it is necessary to use the IC gain history table that has been created offline at DNSC, instead of the gain history table created automatically by OSA. If you have the newest package of IC files from ISDC and these are correctly installed, OSA will find the IC gain history table automatically and you don't need to do anything else. The updated ISDC Instrument Characteristics can be downloaded from http://www.isdc.unige.ch/integral/download/osa_sw However, if you don't have the IC gain table package from ISDC you can download the table from the gain history archive by anonymous ftp through the link provided. Download the table to some suitable subdirectory for your analysis, then set the hidden OSA parameter gainHist: gainHist="/". These files are used instead of the automatically generated gain history tables wherever there has been an unusual or non-linear behaviour of the detector gain. JEM-X1: The Xe line analysis performed using the IC table shows very good gain correction with all Science windows, except the very first science windows and a pair of misfitted outliers, having a Xe level within about 2% of the ideal. Something very odd, neither temperature nor countrate related, happened to cause a sudden gain suppression on this instrument. However most of this problem has been cleared up with the use of the IC tables. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. JEM-X2: The Xe line analysis performed using the IC table shows reasonable gain correction with all except the very first science windows having a Xe level within 4-5% of the ideal. The strange gain suppression in the last 1/3 or so of the revolution is not well-corrected for unit 2 the way it is for unit1, and users should be wary of over-interpreting energy and spectral data in this part of the revolution. As usual it is recommended that users avoid the first few science windows for energy-sensitive applications. CAO 23/03/2014