Gain Notes for Revolution 1150 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. NB: This revolution was affected by solar activity at the beginning of the revolution and it has been suggested that all data up to 4457.5 IJD be classified as falling within a BTI (Bad Time Interval) due to poor gain calibration and other disturbances that may produce unreliable calibration values and poor flux determination. To get optimal results with this revolution it is necessary to use the IC gain history table that has been created offline at DNSC, instead of the gain history table created automatically by OSA. If you have the newest package of IC files from ISDC and these are correctly installed, OSA will find the IC gain history table automatically and you don't need to do anything else. The updated ISDC Instrument Characteristics can be downloaded from http://www.isdc.unige.ch/integral/download/osa_sw However, if you don't have the IC gain table package from ISDC you can download the table from the gain history archive by anonymous ftp through the link provided. Download the table to some suitable subdirectory for your analysis, then set the hidden OSA parameter gainHist: gainHist="/". These files are used instead of the automatically generated gain history tables wherever there has been an unusual or non-linear behaviour of the detector gain. JEM-X1: The Xe line analysis performed using the IC table shows acceptably good gain correction with all except about 6 science windows. All SCWs in the second half of the revolution are within 1-2% of the ideal, with most of the other SCWs having a Xe level within 3% of the ideal. A handful of SCWs within the BTI are are only good to 4-5%. There is a great deal of scatter on the data that cannot be corrected especially in the first 1/3rd of the revolution, as well as several misfitted outliers. Outliers are all SCWs with a fitted Xe line width smaller than 2 and greater than 4. As usual, it is recommended that users avoid the first few science windows for energy-sensitive applications. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. JEM-X2: The Xe line analysis performed using the IC table shows very good gain correction with all except the very first science windows having a Xe level within -32% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. CAO 23/03/2012