Gain Notes for Revolution 1113 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. NB: The end of this revolution was strongly affected by solar activity, so that from IJD 4347.7 onward, the revolution is declared a BTI, and standard OSA processing will ignore all science data after this time. Though results with the IC table show fine energy calibration during this period it is certain that other quanitities like the flux will not be good, and spectra from this BTI should not be used. To get optimal results with this revolution it is necessary to use the IC gain history table that has been created offline at DNSC, instead of the gain history table created automatically by OSA. If you have the newest package of IC files from ISDC and these are correctly installed, OSA will find the IC gain history table automatically and you don't need to do anything else. The updated ISDC Instrument Characteristics can be downloaded from http://www.isdc.unige.ch/integral/download/osa_sw However, if you don't have the IC gain table package from ISDC you can download the table from the gain history archive by anonymous ftp through the link provided. Download the table to some suitable subdirectory for your analysis, then set the hidden OSA parameter gainHist: gainHist="/". These files are used instead of the automatically generated gain history tables wherever there has been an unusual or non-linear behaviour of the detector gain. JEM-X1: The Xe line analysis performed using the IC table shows excellent gain correction with all except a couple of science windows having a Xe level within 2% of the ideal, and the others within 3%. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. JEM-X2: The Xe line analysis performed using the IC table shows excellent gain correction with all except two science windows having a Xe level within 2% of the ideal, the outstanding pair are within 4% of the ideal. All science data beyond IJD of 4347.7 will be ignored by OSA processing since the period beyond this time, and into the first part of revolution 1114 has been a BTI. As usual however, it is also recommended that users avoid the first few science windows in this revolution for energy-sensitive applications. CAO 1/03/2012