Gain Notes for Revolution 1102 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. NB: This is another revolution disrupted by solar activity with significant gain suppression on both unit microstrip plates, and considerable differential gain variation (differences in gain behaviour between the heavily irradiated calibration spectra plate areas and the lightly irradiated science data areas), leading to large variations in the energy correction during the revolution. The IC tables for each unit have been able to correct for most of this unwanted gain variation, but a +/- 5-6% scatter in the data still remains. This is unprecedented in the mission, but cannot be corrected for at the moment. If better IC correction can be achieved new tables will be delivered to ISDC. To get optimal results with this revolution it is necessary to use the IC gain history table that has been created offline at DNSC, instead of the gain history table created automatically by OSA. If you have the newest package of IC files from ISDC and these are correctly installed, OSA will find the IC gain history table automatically and you don't need to do anything else. The updated ISDC Instrument Characteristics can be downloaded from http://www.isdc.unige.ch/integral/download/osa_sw However, if you don't have the IC gain table package from ISDC you can download the table from the gain history archive by anonymous ftp through the link provided. Download the table to some suitable subdirectory for your analysis, then set the hidden OSA parameter gainHist: gainHist="/". These files are used instead of the automatically generated gain history tables wherever there has been an unusual or non-linear behaviour of the detector gain. JEM-X1: The Xe line analysis performed using the IC table shows poor gain correction with all science windows having a Xe level within5-6% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. JEM-X2: The Xe line analysis performed using the IC table shows poor gain correction with all science windows having a Xe level within 5-6% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. CAO 06/12/2011