Gain Notes for Revolution 1018 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination a IC Gain History table is needed. JEM-X1: The Xe line analysis performed using the IC table shows very gain correction with all except the very first science windows, and a few misfitted outliers, having a Xe level within 1-2% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. JEM-X2: The Xe line analysis performed using the IC table shows excellent gain correction with all except the very first science windows, and the last one, having a Xe level within 1% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. CAO 22/02/2011