Gain Notes for Revolution 783 JEM-X1: A quite normal revolution with fine gain fitting. All the data from this revolution should be useable. The Xe analysis confirms that the energy determination for this revolution is okay, within about 3% of the ideal. However, as for all revolutions, the first few science windows should be avoided in any analysis requiring very good gain determination. Since the temperature variation of the gain has tripled since the beginning of the mission, future version of OSA (version 8.0), including temperature dependency, will benefit this revolution when it becomes available in 2009. This is the first full revolution after the loss of the fourth calibration source, but this is no longer a problem since the science data is now corrected without the use of this source. JEM-X2: No scientific data or gain history data for this unit. CAO 23/3/2009