Gain Notes for Revolution 727 JEM-X1: A quite normal revolution with fine gain fitting. All the data from this revolution should be useable. The Xe analysis confirms that the energy determination for this revolution is okay, within about 3% of the ideal. However, as for all revolutions, the first few science windows should be avoided in any analysis requiring very good gain determination. Since the temperature variation of the gain has tripled since the beginning of the mission, future version of OSA (version 8.0), including temperature dependency, will benefit this revolution when it becomes available in 2009. Also, the general level of the Xe line appears to be quite low. This may have to do with the fact that the gain of the instrument is currently very high and this puts the Xe peak on a sharp falloff at high energy that surpresses the peak on its high side and hence pushes the position of the peak to lower energies. If this is the case, no correction for this low Xe line needs to be performed since all the lower energy photons will still have been corrected to their proper value. JEM-X2: No scientific data or gain history data for this unit. CAO 31/10/2008